  1.         COMPUTERIZED VIDEOKERATOSCOPES DATA DISPLAY
  2.  
  3. Curvature Maps
  4.  
  5. The values used to draw the curvature maps 
  6. can be calculated based on the axial or the 
  7. instantaneous corneal radius.
  8.  
  9. The axial radius (also called the sagittal 
  10. radius) is the distance normal to a 
  11. peripheral corneal point to a single axis.  
  12.  
  13. The instantaneous radius (also called 
  14. tangential, true or local radius) is a radius 
  15. of a peripheral point which is independent 
  16. of any axis of the cornea.  With the cornea, 
  17. as it becomes more aspheric in the 
  18. periphery, the center of curvature moves 
  19. away from the central axis of the cornea 
  20. forming a locus of centers of curvatures.  
  21. The instantaneous radius is more accurate and sensitive to 
  22. changes in curvature.  It is longer than the axial radius.